Sign In | Join Free | My esadidasol.com |
|
Model Number : H2/C-5000
Place of Origin : Switzerland
MOQ : 1PCS
Price : Negotiable
Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability : 5000PCS
Delivery Time : 5-8 work days
Packaging Details : Standard packaging
Nominal Range : 0 - 5000 ppm
Maximum Overload : 10000 ppm
Output Signal : 9.5 ± 2.5 nA/ppm
Resolution (Electronics dependent) : < 5.3 ppm
T90 Response Time : < 45 s
Typical Baseline Range (pure air, 20°C) : -40 ppm to 30 ppm
H2/C-5000 3-Electrode Electrochemical Hydrogen Gas Sensor for Safety and Process Control
Applications:
• Safety and Process Control
Important Application Notes:
H2 sensors lose >10% of their initial sensitivity within the first month after production before the output stabilizes.
Rec. Load Resistor | 10 - 33 Ω |
Bias (V_Sens-V_Ref) | not recommended |
Conformity to RoHS directive | RoHS Compliance |
Relative Humidity Range | 15 % to 90 % RH non-condensing |
Temperature Range | -40 °C to 50 °C |
Pressure Range | Atmospheric ± 10% |
Pressure Coefficient | N.D. |
Humidity Effect | None |
Expected Operation Life | 2 years in air |
Expected Long Term Output Drift in air | < 2 % signal loss per month after stabilization |
Filter Life | 6 months in container |
Rec. Storage Temperature | 5°C - 20°C |
Warranty Period | 12 months from date of dispatch |
![]() |
H2/C-5000 3-Electrode Electrochemical Hydrogen Gas Sensor for Safety and Process Control Images |